T&H INSTRUMENT CO.,LTD
Hotline
0919203336

D350A_V5C40AT8

DETAILS

 

 

 

 

 

 

 

 

 

 

 

D350A_V5C40AT8 PERFORMANCE:

ITEM

INSTRUMENT PERFORMANCE

RANGE

GEAR

RANGE

ACCURACY(SET,READ)

RESOLUTION(SET,READ)

Voltage1

5V (DC to -5V)*1

±0.01%RD±0.01%FS

≤50uV,  ≤10uV

Current1

10mA

±0.01%RD±0.01%FS

≤100nA, ≤20nA

Current 2

100mA

±0.01%RD±0.01%FS

≤1uA, ≤200nA

Current 3

1A

±0.01%RD±0.01%FS

≤10uA, ≤2uA

Current 4

12A

±0.01%RD±0.01%FS

≤120uA, ≤24uA

Current 5

40A

±0.01%RD±0.01%FS

≤400uA, ≤80uA

Current set range

1uA ~ 40A (MAX320A In parallel mode)

No. of CHANNELs

 

8Chl

No. of STEPs

 

1024

TIME

Accuracy

≤±5uS/S (Equipment autonomous control, no communication and other time delay errors)

Min Step Time

20mS

Min Pulse Time

2mS

Sampling Interval

≤10mS*2

INPUT IMPEDENCE

 

≥10GΩ(INPUT BASE CURRENT < 2.0nA)

TEMPERATURE DRIFT

 

≤20ppm/℃

RESPONSE TIME

 

Current 0~90% ≤1mS; Current -90~90% ≤3mS

DATA RECORD

Sampling Rate

100Hz(Sampling interval 10mS)*2

Setting Method

Interval time, Sampling rate, Interval current, Interval capacity, Interval voltage, Interval energy

PROCESS CONTROL

Work Mode

HOLD / CC / CV / CCCV / CRATE / CP / CR / CPULSE / CSLOPE / DC / DV / DCDV / DRATE / DP / DR / DPULSE / DSLOPE

End Conditions

TIME / VOLTAGE / CURRENT / POWER / CAPACITY / ENERGY / LOOPS / PHASE CYCLE

Protective Function

Power-off data protection, Net break protection

User-defined safety: Over voltage, Under voltage, Over current, Over rate, Over capacity.

FUNTION

 

DCIR、Capacitance capacity calculation、Step pluse、Three-electrode test(Reference test)、DQ/DV

OUTPUT

 

Four electrodes (support reference electrode test)

COMMUNICATION

 

10/100M Ethernet(RJ45)、USB(Console)

OPERATING ENVIRONMENT

Temperature

0~40℃(23±2℃Optimum accuracy)

Humidity

≤90% No condensation

POWER SUPPLY

 

AC220V~±10% 50Hz / AC110V~±10% 60Hz,  2600W 3

EQUIPMENT SIZE

 

6U  W*H*L:430*540*178(mm)  32KG

PRODUCTS

RELATED PRODUCTS
Support online