T&H INSTRUMENT CO.,LTD
Hotline
0919203336

M340A_V5C100MT8

DETAILS

 

 

 

 

 

 

 

MODEL:  M340A_V5C100MT8

PERFORMANCE:

ITEM

INSTRUMENT PERFORMANCE

 

RANGE

GEAR

RANGE

ACCURACY(SET,READ)

RESOLUTION(SET,READ)

Voltage1

5V (DC to -5V)*1

±0.01%RD±0.01%FS

≤50uV,  ≤10uV

Current1

100uA

±0.01%RD±0.01%FS

≤1nA,   ≤200pA

Current 2

1mA

±0.01%RD±0.01%FS

≤10nA,  ≤2nA

Current 3

10mA

±0.01%RD±0.01%FS

≤100nA, ≤20nA

Current 4

100mA

±0.01%RD±0.01%FS

≤1uA,   ≤200nA

Current set range

10nA ~ 100mA

No. of CHANNELs

 

8Chl

No. of STEPs

 

1024

TIME

Accuracy

≤±5uS/S (Equipment autonomous control, no communication and other time delay errors)

Min Step Time

20mS

Min Pulse Time

2mS

Sampling Interval

≤10mS*2

INPUT IMPEDENCE

 

≥10GΩ(INPUT BASE CURRENT < 0.5nA)

TEMPERATURE DRIFT

 

≤20ppm/℃

RESPONSE TIME

 

Current 0~90% ≤1mS; Current -90~90% ≤3mS

DATA RECORD

Sampling Rate

100Hz(Sampling interval 10mS)*2

Setting Method

Interval time, Sampling rate, Interval current, Interval capacity, Interval voltage, Interval energy

PROCESS CONTROL

Work Mode

REST / CC / CV / CCCV / CRATE / CP / CR / CPULSE / CSLOPE / DC / DV / DCDV / DRATE / DP / DR / DPULSE / DSLOPE

End Conditions

TIME / VOLTAGE / CURRENT / POWER / CAPACITY / ENERGY / LOOPS / PHASE CYCLE

Protective Function

Power-off data protection, Net break protection

User-defined safety: Over voltage, Under voltage, Over current, Over rate, Over capacity.

FUNTION

 

DCIR、Capacitance capacity calculation、Step pluse、Three-electrode test(Reference test)、DQ/DV

OUTPUT

 

Four electrodes (support reference electrode test)

COMMUNICATION

 

10/100M Ethernet(RJ45)、USB(Console)

OPERATING ENVIRONMENT

Temperature

0~40℃(23±2℃Optimum accuracy)

Humidity

≤90% No condensation

POWER SUPPLY

 

AC220V~±10% 50Hz / AC110V~±10% 60Hz,  40W 3

EQUIPMENT SIZE

 

1U  W*H*L:430*300*43.5(mm)

PRODUCTS

RELATED PRODUCTS
Support online